Digital Fine Differential Pressure Sensor Model GTI 135 & GTI 145

● Switch function (dual NPN or PNP open collector output)
● Analog output 1~5VDC or 40~20mA, also can be customized
● Differential pressure display, analog output, digital filter, key lock, peak and valley hold display and zero point adjustment
● Easy to install, wall-mounted or bracket-mounted two installation forms are optional
● 30x30mm and 24x48mm two sizes, to meet different options

  • GTI
  • China
  • One week
  • Strong

Details

Two compact differential pressure digital display meters use high-voltage resistant micro-silicon resistance sensors and advanced digital technology to achieve highly sensitive and reliable differential pressure testing. 


Specifications


Differential Range

Testing Accuracy

Resolution (Pa)

Overload

GTI135

GTI145

0~50Pa

±50Pa

±(2%F.S.+1digit)@23

0.1

 

 

10kpa

0~100Pa

±100Pa

0~200Pa

±200Pa

±(1.5%F.S.+1digit)@23

 

1

0~500Pa

±500Pa

0~1kPa

±1kPa

±(1%F.S.+1digit)@23

 

20kpa

0~2kPa

±2kPa

0~5kPa

±5kPa

10

1